Krouwer Consulting
 

 

Home
Up
Modeling
Freq. of Med. Errors
Frequency of Medical Errors II
Six Sigma
Six Sigma II
Six Sigma III
Outliers
Equivalent QC
Equivalent QC2
Sys. not People
Near Miss
Detection
Recovery
Risk Management
FMEA FRACAS
Fault Tree Example
Beware the Technical Administrator
Discrepant Analysis
FMEA Validation
Proficiency Testing and Six Sigma
It's Up To the Lab Director
Bland Altman Plots
Risk Management II
Unit Use QC
The reverse Pareto – not a good
Risk Management III
Pedigree of Approaches to Reduce Error
What it Takes to Get New Ideas Accepted
FMEA FRACAS FTs Pareto
When to conduct FMEA
Error Reporting Systems for Clin
Customer Misuse
Fault Isolation
Uniformity of Claims
Outliers in QC and Proficiency Testing
The quality of quality initiativ
Publishing in journals or on web
FDA Waiver Guidance
Pay for Performance
Pay for performance II
Performance vs. Attribute Goals
ISO Terminology
Latent Errors
Preventable medical errors
Troponin I
GUM Comment
More on GUM
Standards Issues
Slow Qual. Progress
More QC?
Report Writing
Software Validation
Dump Data

Some technical comments about Six Sigma in the lab

 

Everyone’s talking about six sigma these days.

 

To recall, six sigma implies a process with 3.4 defects per million. Assume that one has a set of measurements that are normally distributed, for example lab glucose differences from reference. Many people remember that ± 2 standard deviations includes 95% of the values. The idea behind six sigma is that if one sets goals of ± 6 standard deviations, there will be very few defects (where defects are defined as being outside of the ± 6 standard deviation limits = 6 sigma. Of course, these limits must be clinically meaningful, one can’t just multiply the observed standard deviation by six to get the limits. So one starts with limits and sees where the measured standard deviation is and calculates how many “sigma” the process exhibits.

 

How do you get 3.4 defects – When I go through the math (1) to get the number of defects for a six sigma process, I get 0.00197, not 3.4 defects:

 

Defects per million = 1,000,000 * number of defects

where number of defects = (1 – probability good)

where probability good = F(6 sigma) – F(-6 sigma)

where F() = the cumulative distribution function of the normal distribution

 

What’s wrong? Six sigma assumes a 1.5 sigma bias, so if one repeats the calculations where probability good = F(1.5 + 6 sigma) – F(1.5 - 6 sigma), one gets 3.4.

 

Six sigma is defined for a normally distributed process. – Well, one might make the case that lab glucose differences from reference might be normally distributed but then again they might not, especially since when things go wrong there are outliers which tend to make things non normal. If the data are non normal, the six sigma numbers no longer apply.

 

But what does six sigma mean if one is talking about attribute data such as patient ID problems rather than continuous variables such as glucose differences. A patient ID mix up is clearly a defect but this does not come from a normally distributed process. Each attempt to match a patient ID and sample is either correct or incorrect. While one can count defects per million, this has nothing to do with “sigma” (e.g., a standard deviation).

 

A six sigma process has bias – Hard to believe but a six sigma process is defined to have a bias of 1.5 sigma. So what the big deal with this bias, if inclusion of a 1.5 sigma bias for a six sigma process still results in only 3.4 defects per million. The problem has been explained by George Klee, who shows that even a small bias can lead to increased patient misclassifications (2), which raises the risk for a wrong patient treatment. As an aside, a six sigma process with its 1.5 sigma bias, would not be suitable for the ISO standard GUM (the guide to the expression of uncertainty in measurement) since GUM requires all biases to be eliminated. So where did this bias come from. The originators of the six sigma concept were from Motorola, where it was observed that the average process tended to have a 1.5 sigma (drift) bias. Hence the inclusion of this bias was to be representative of real processes.

 

Does six sigma help labs, hospitals, or diagnostic companies – The above comments have nothing to do with the desirability of usefulness of six sigma. They are technical comments / answers for questions that I had. Of course no one will build in a bias to try to achieve six sigma! Six sigma is really a collection of quality improvement tools (3) and as such the name is somewhat misleading. A better name might be Total Quality Management – II (or a higher number). For example, FMEA (Failure Mode Effects Analysis) is part of Six Sigma and may involve an entire program without the measurement of a single standard deviation. Most “black belts” are managers and are neither statisticians nor people who have a career in data analysis and quality. The success of six sigma in organizations is a function of management commitment and training.

 

References

 

  1. Lucas JL. The essential six sigma Quality Progress 2002;35:27-31.

 

  1. Klee GG. Analytic performance goals based on direct effect of analytic bias on medical classification decisions. CDC 1995 Institute: Frontiers in Laboratory Practice Research. pp 219-226. Available online at : www.phppo.cdc.gov/dls/pdf/institute/klee.pdf

 

  1. Hahn GJ, Hill WJ, Hoerl RW, and Zinkgraf SA. The impact of six sigma improvement – A glimpse into the future of statistics. The American Statistician 1999;53:208-215